black piece in banner
Applications News Investors
 

Klarite SERS Peformance

Reproducibility - no hot spots

Reproducibility data from 567 spectra taken at 10um intervals over a 200um x 260um area on Klarite SERS substrate. Magnitude of the 1072cm-1 Raman peak is shown. Average signal level 30,000 counts. Relative standard deviation observed <7%.

Spectra recorded on Renishaw’s inVia Reflex Raman microscope with 785nm excitation, 10mW laser power, 1sec exposure, 50x magnification, confocal configuration. Sample Benzenethiol monolayer

Glucose

Spectra recorded on Renishaw’s inVia Reflex Raman microscope with 785nm excitation, 10mW laser power, 10sec exposure time. No background subtraction has been applied.

Preliminary Klarite SERS spectrum of 0.1 Molar Glucose solution.

Histamine

Spectra recorded on Renishaw’s inVia Reflex Raman microscope with 785nm excitation, 10mW laser power, 10sec exposure time. No background subtraction has been applied

Typical Klarite SERS spectrum of 0.075 Molar partially crystalline Histamine.

Urea

Spectra recorded on Renishaw’s inVia Reflex Raman microscope with 785nm excitation, 10mW laser power, 10sec exposure time. No background subtraction has been applied.

Typical Klarite SERS spectrum of 2.3x10-4 Molar Urea solution dehydrated to form non-crystalline thin film.

Phenylalanine

Spectra recorded on Renishaw’s inVia Reflex Raman microscope with 785nm excitation, 10mW laser power, 10sec exposure time. No background subtraction has been applied.

Preliminary Klarite SERS spectrum of 0.01 Molar Phenylalanine at different points of crystallisation on dehydrated sample.

Benzenethiol

Typical surface enhanced Raman spectrum obtained with a Benzenethiol sample on a Klarite substrate.

Sample- Benzenethiol, laser power at 633 and 785nm <5mW, integration time 10 seconds, Raman systems Renishaw RM1000 and Renishaw RM2000